PE500DF and PE500DUV
Operational Specifications
Description
Nominal
Range
Power
500 watts
340-575 watts
Current
27 amps (DC)
20-32 amps (DC)
Operating Voltage
18.5 volts (DC)
16-21 volts (DC)
Ignition Voltage
35 kilovolts (recommended minimum)
Temperature
150º C (maximum)
Lifetime*
1000 hours (500 hour warranty)
* End of lamp life is defined as 50% of initial output.
Output at Nominal Power
F= UV Filtered Output/ UV = UV Enhanced Output
Description
Peak Intensity
Radiant Output*
UV Output*
IR Output*
Visible Output*
Color Temperature
Peak Instabilities
Beam Geometry**
PE500DF
24x105 candelas
112 watts
5.5 watts
65 watts
10500 Lumens
5600 Kelvin
4%
3º/4º/5º
PE500DUV
17x105 candelas
112 watts
11.5 watts
62 watts
9600 Lumens
5050 Kelvin
4%
3º/4º/5º
*These values indicate total output in all directions.
Wavelengths = UV<390nm, IR>770nm, Visible 390nm to 770nm.
** Beam Geometry defined as half angle at 10% PTS after 01/100/1000 hours.
Dimensions
.22
(5.59)
2.00 (50.8)
.04 (1.02)
.250 (6.35)
2.00 (50.8)
2 x .60 (15.2)
8-32 UNC 2B x .25 DP
3 Holes EQ SP on A
1.60 (40.64) BC
.273 (6.93)
2.858
(72.59)
.066 (1.52)
.20 (5.08)
2.592 (65.84)
Y
X
Focal Point
Cathode
Dimensions in parentheses are in millimeters.
Physical Specifications
Description
Arc Gap
Reflector Geometry
Weight
Window Diameter
Specification
.09 inch (2.28mm)
Parabolic Y2 = 0.8 X (inch)
664 grams
2 inch (50.8 mm)
Focused Output at F/1.0 Lens
Description
Visible Output
6mm aperture
5200 Lumens
8mm aperture
2300 Lumens
* Nominal values at 500 watts after 2 hour burn-in.
Total Output*
48 watts
22 watts
Notes
1. Lamp must not be operated with window facing upwards within
45º of vertical.
2. Seal temperature must not exceed 150º C.
3. Current/power regulated power supplies and PerkinElmer lamp
housing units are recommended.
4. Lamp must be operated within recommended current and power
range. Over powering may lead to arc instability, hard starting
and premature aging.
5. CERMAX lamps are much safer lamps to use than their quartz
xenon arc lamp equivalents. However, caution must be practiced
when operating lamps because they are under high pressure,
require high voltage, reach temperatures up to 200º C, and their
IR and UV radiation can cause skin burns and eye damage.
Read hazard sheet included with each lamp shipment.
Spectral Output
F
UV
F + HOT MIRROR
100
80
60
40
20
0
200
300 400 500
600 700
800
900 1000 1100
WAVELENGTH (NM)
PerkinElmer welcomes inquiries about special lamps. We would be pleased to discuss the requirements of your application and the feasibility of
designing a lamp specifically suited to your needs.
For more information e-mail us at opto@perkinelmer.com or visit our web site at www.perkinelmer.com/opto.
All values are nominal; specifications subject to change without notice.
USA:
Europe:
Asia:
PerkinElmer Optoelectronics
PerkinElmer Optoelectronics GmbH
PerkinElmer Optoelectronics
44370 Christy St.
Wenzel-Jaksch-Str. 31
47 Ayer Rajah Crescent #06-12
Fremont, CA 94538
65199 Wiesbaden
Singapore 139947
Phone: (510) 979-6500
Germany
Phone: +65 775 2022
Fax: (510) 687-1140
Phone: +49 611 492 0
Fax: +65 775 1008
.
Toll Free: (800) 775-6786
Fax: +49 611 492 369
© 2000 PerkinElmer inc. All rights reserved.
www.perkinelmer.com/opto
DS-157 Rev C 1101

Share Link: 

datasheetbank.com [ Privacy Policy ] [ Request Datasheet ] [ Contact Us ]