Measurement capabilities
3dB cut-off frequency (S21),
Responsivity (S21),
Electrical reflection (S11 or S22),
Group Delay vs. frequency,
Insertion Loss (IL),
Transmission bandwidth,
all electrical S-parameter measurements.
Target test devices
Transmitter (E/O)
0DFK=HKQGHUPRGXODWRUV
(OHFWURDEVRUSWLRQPRGXODWRUV($0
'LUHFWO\PRGXODWHGODVHUV
7UDQVPLWWHURSWLFDOVXEDVVHPEOLHV726$
Receiver (O/E)
PIN diodes
Avalanche photodiodes (APD)
Receiver optical subassemblies (ROSA)
Optical (O/O)
Passive optical components
Optical fibers and filters
Optical transmission systems
Agilent N4374B Specifications
Measurement conditions
Modulation frequency range from 10 MHz to 4.5 GHz
Foreward and reverse RF power +5 dBm
Number of points 899
Number of averages: 1
IFBW 300 Hz
Network analyzer set to “stepped sweep – sweep
moves in discrete steps”
After full two-port electrical calibration using an
Electronic Calibration Module, Agilent 85092C,
at constant temperature (±1° C)
Modulator bias optimization set to “every sweep”
Measurement frequency grid equals electrical
calibration grid
DUT signal delay ≤ 0.1/IF-BW
Specified temperature range: +20° C to +26° C.
After warm-up time of 90 minutes
Using high quality electrical and optical connectors
and RF cables in perfect condition
Using supplied RF cables (8120-8862)