First Prev 11 12 13 14 15 16 17 18 19 20 Next Last
Test port output power1
Table 12. Option 230/235/240/245/260/265/280/285/430/435/440/
445/460/465/480/485
Description
Nominal power
(preset power)
Level accuracy2, 6
(stepped sweep mode)
Level accuracy2
(swept sweep mode)
Level linearity3, 5, 6
(stepped sweep mode)
9 kHz to 5 GHz
5 GHz to 6 GHz
6 GHz to 7 GHz
7 GHz to 8.5 GHz
Level linearity5
(swept sweep mode)4
9 kHz to 5 GHz
5 GHz to 6 GHz
6 GHz to 7 GHz
7 GHz to 8.5 GHz
Range5, 6
9 kHz to 5 GHz
5 GHz to 6 GHz
6 GHz to 7 GHz
7 GHz to 8.5 GHz
Sweep range5, 6
9 kHz to 5 GHz
5 GHz to 6 GHz
6 GHz to 7 GHz
7 GHz to 8.5 GHz
Level resolution
Specification
0 dBm
±0.650 dB (at 0 dBm,
50 MHz absolute)
±1.0 dB
Typical
±2.5 dB
±0.75 dB (–20 to 10 dBm)
±0.75 dB (–20 to 9 dBm )
±0.75 dB (–20 to 8 dBm )
±0.75 dB (–20 to 7 dBm )
–55 to 10 dBm
–55 to 9 dBm
–55 to 8 dBm
–55 to 7 dBm
–55 to 10 dBm
–55 to 9 dBm
–55 to 8 dBm
–55 to 7 dBm
0.05 dB
±1.5 dB (at –20 to 10 dBm)
±1.5 dB (at –20 to 9 dBm)
±1.5 dB (at –20 to 8 dBm)
±1.5 dB (at –20 to 7 dBm)
Description
Max leveled power5, 6
(Option 230/235/240/245/260/
265/280/285)
9 kHz to 8.5 GHz
(Option 430/435/440/445/460/
465/480/485)
9 kHz to 6 GHz
6 GHz to 7 GHz
7 GHz to 8.5 GHz
Specification
SPD
10 dBm
10 dBm
9 dBm
8 dBm
1. Source output performance on port 1 only. Other port output performance is typical.
2. Level accuracy is taken at 0 dBm, relative to 50 MHz reference unless otherwise stated.
3. Level linearity given is relative to 0 dBm.
4. The specification might not be met at 5 MHz or 50 MHz.
5. The level accuracy specification needs to be taken into account for test port output power level.
6. Power calibration using an external power meter improves level accuracy of the test port output power.
Proper power meters/sensors, and the 82357B USB-GPIB interface are required to conduct power calibration.