High-Temperature Logger iButton with 8KB
Data-Log Memory
ADDRESS
0219h
021Ah
BIT 7
0
0
BIT 6 BIT 5 BIT 4
10 Seconds
10 Minutes
BIT 3
BIT 2 BIT 1
Single Seconds
Single Minutes
BIT 0 FUNCTION ACCESS*
021Bh
021Ch
0
12/24
20 Hour
AM/PM
10 Hour
0
0
10 Date
Single Hours
Single Date
Mission
Timestamp
R
R
021Dh CENT
0
0
10
Months
Single Months
021Eh
021Fh
10 Years
Single Years
(No Function; Reads 00h)
—
R
R
0220h
0221h
0222h
0223h
0224h
0225h
0226h
0227h
0228h
…
022Fh
0230h
…
0237h
0238h
…
023Fh
Low Byte
Center Byte
High Byte
Low Byte
Center Byte
High Byte
Configuration Code
EPW
First Byte
…
Eighth Byte
First Byte
…
Eighth Byte
(No Function; All These Bytes Read 00h)
Mission
Samples
R
R
Counter
Device
Samples
R
R
Counter
Flavor
R
R
PW Control R/W R
Read
Access
W—
Password
Full
Access
W—
Password
—
R
R
*The left entry in the ACCESS column is valid between missions. The right entry shows the applicable access type while a
mission is in progress.
Figure 6. DS1922E Register Pages Map (continued)