how it affects measurement accuracy as temperature changes can be
made. Once the amount of drift is known, you can determine if a new cal-
ibration is required. A good rule of thumb is to recalibrate the converter
(or system) with every ten degrees of ambient temperature change.
11) What do the numbers in the calibration registers actually mean?
A: There are two internal read/write calibration registers in the CS5525/26
(offset, and gain). One LSB in the offset register is 2-24 proportion of the
input span (bipolar span is 2 times the unipolar span). The MSB in the
offset register determines if the offset that is to be trimmed is either posi-
tive or negative. The converter can typically trim ±50% of the input span.
The gain register spans from 0 to (2 - 2-23). The decimal equivalent
meaning of the gain register is:
D = b020+ b12-1 + b22-2 + b32-3 + ... + bN2-N
where the binary numbers have a value of either zero or one. After a gain
calibration has been performed, the numeric value in the gain register
should not exceed the range of 0.5 to 2.0 (decimal) [400000(Hex) to
FFFFFF(Hex)].
12) How can the gain be calibrated if a full-scale signal is not available?
A: The CS5525/26 can be gain calibrated with some input signal other than
full scale. For example, when the converter is reset, the gain register’s
calibration word is 1.0 (decimal). If a signal representing ten percent of
full scale reads three percent less than it should, the value in the gain reg-
ister can be scaled up by three percent. Gain accuracy can be improved if
output words are averaged while using this technique. Use caution when
a calibration signal less than full scale is being used. If the transfer func-
tion of the transducer being used to generate the ten percent signal hap-
pens to have a major nonlinearity at the point at which calibration is
being performed, it will cause the rest of the transfer function to be incor-
rect.