Figure 4 illustrates the CS5525/26 connected to
measure ground referenced unipolar signals of a
positive polarity using the 1 V, 2.5 V, and 5 V input
voltage ranges on the converter. For the 25 mV, 55
mV, and 100 mV ranges the signal must have a
common mode near +2.5 V (NBV = 0V).
The CS5525/26 are optimized for the measurement
of thermocouple outputs, but they are also well
suited for the measurement of ratiometric bridge
transducer outputs. Figure 5 illustrates the
CS5525/26 connected to measure the output of a
ratiometric differential bridge transducer while op-
erating from a single +5 V supply.
2N 5087
or sim ilar
NBV
10µF
+
34.8K
30.1K
2.0K
NBV
2.1K
+
10 µF
Figure 2. Charge Pump Drive Circuit for VD+ = 3 V.
-5V
-5V
Figure 3. Alternate NBV Circuits.
+5V
A nalo g
S upply
10
0 .1 µF
0 .1 µF
2.5V
2
VA+
20
VREF+
13
VD+
10
XOUT
500k
20 pF
19 VREF-
0 to +5V Input
+
-
CM = 0 to VA+
C S5525 XIN 9
3
CS5526
AIN +
4
A IN -
1 AGND
16 A3
15 A2
7 A1
6 A0
CS 18
SCLK 11
SDI 17
14
SDO
NBV CPD DGND
32.768 ~ 100kH z
O ptional
C loc k
S o u rc e
S e rial
D ata
Interface
5
8
12
Figure 4. CS5525/26 Configured for ground-referenced Unipolar Signals.
DS202F3
9